Driving the future of technology : Original solution in measurement and control

 Lecture for Measurement

Posted 2019/9/23

Current amplifier (Transimpedance amplifier)

To measure a very small current signal

2. Noise reduction method after signal amplification

There are two main methods to reduce voltage noise after signal amplification below.

  •  Setting appropriate measurement band
  •  S/N ratio improvement by averaging

Setting appropriate measurement band

In order to reduce noise, it is effective to set an appropriate measurement band, that is, to remove noise components in unnecessary frequency bands. In particular, even if the noise amplitude level is the same, the noise at higher frequency has larger energy, so it is important to remove unnecessary high-frequency noise component.

The output waveforms with and without a low-pass filter (LPF) (1-pole, fc = 1 kHz) to the output of the current amplifier are shown in Figure 11. By using LPF, it is seen that the noise is reduces and the output waveform become very sharp. In this way, by setting only the signal of the band to be measured by a technique such as filtering, S/N ratio can be improved and a smaller current signal can be detected. In addition, as an actual measurement example, a 1-pAp-p real-time signal was successfully detected using SA-607F2.

Figure 11: Comparison of output waveforms due to difference in measurement bandwidth

(a) Block diagram of measurement

 

(b) Output waveform without LPF (BW: 20 kHz)

 

(c) Output waveform with LPF (BW: 1 kHz)

 

Guidelines of the measurement band settings for signal waveform measurement are shown in Table 6.
Note that Table 6 only shows guidelines, and the appropriate measurement band may varies depending on the customer’s measurement content. For that reason, various filters are provided.

 

Table 6: Guidelines for setting the measurement band suitable for the signal waveform to be measured (when using LPF)

Measured signal waveform Cutoff frequency of LPF
Sine wave 3–5 times higher than the signal frequency
Square wave About 10 times higher than the signal frequency
Triangle wave 7–10 times higher than the signal frequency

 

Table 7: Various filter products

Type Overview Product
Cutoff frequency that can be set freely
  •  Multifunction filter
  •  Dual-channel programmable filter
  •  High-frequency continuous variable filter
Multichannel analog-signal-preprocessing system that makes it possible to configure filters, differential amplifiers, isolation amplifiers, etc. according to the required number of channels
  •  Differential amplifier
  •  Isolation amplifier
  •  Multifunction filter
  •  GPIB unit
Board-mounted type filter module
  •  Resistor-tuning filter
  •  Voltage-tuning filter
  •  Programmable filter
  •  Fixed-frequency filter
Power-supply storage case with built-in filter module A storage case that can built-in various filter modules is available.

Improvement of S/N ratio by averaging

Even with the proper measurement system and measurement conditions, a small signal from the sensor may be buried in noise and not observed in real time measurement. Even in this case, it may be possible to detect small signals by averaging. The conditions under which small signals can be detected by averaging are as follows.

  • (a) The noise component is random noise.
  • (b) The signal want to detect is a repeating waveform.
  • (c) The signal want to detect has the synchronization signal.

Block diagrams of the measurement system and the output waveforms of the current amplifier with and without averaging are shown in Figure 12. It can be seen that the averaging enables detection of very small current signals buried in noise.
In averaging, it is important to set the measuring range of the measurement instrument appropriately. Then let’s discuss how to utilize the averaging function of oscilloscope. When the signal exceeds the voltage measurement range of oscilloscope, the part of the waveform that exceeds the measurement range is clamped, so signals cannot be detected accurately. Also, if the voltage measurement range is too wide and the detected signal level is under the measurement resolution, the signal cannot be detected regardless of reduction of random noise by averaging. Therefore, it is important that dynamic range setting of the measurement instrument for the S/N ratio improvement by the averaging.

In this way, it is important to set the appropriate measurement instrument and measurement range in consideration of the signal level to be measured and the noise level superimposed on the signal.

Figure 12: Comparison of output waveforms with and without by averaging

(a) Block diagram of measurement

 

(b) Output waveform without averaging

 

(c) Output waveform with 512-times averaging

Related Product

Wideband Current Amplifier SA-600 Series

World's top level current amplifier with extremely low noise

     Wide current range measurement (sub pico-Amp. to 1 µA)
  •  Realize both high gain and wide bandwidth
  •  Low noise current density (fA/√Hz order)

 


Transimpedance amplifier (current amplifier) IV-204F3 and IV-202F4

Compact, Wideband current amplifier

     I/V conversion of high-frequency signals up to 10 MHz
  •  High frequency and low noise are balanced at high level.

 


Wideband Current Amplifier CA-550 Series / CA-650 Series

CA-550 series and CA-650 series are low noise over wideband current amplifiers with a high gain and wide bandwidth.

 

Related keywords :  Ultra low noise amplifier, LNA

back to the page top